ADM Analog Digital ModuleSysTec ADM Analog Digital Module (A/D Converter) for WT1, WT3, WT4000, WT6000 and WT8000 Terminals
- Description
- Additional Information
- 01 / 05SysTec ADM scale module
The analog load cell input of a SysTec terminal.
The SysTec ADM (Analog Digital Module, SKU 20495) is the scale interface that connects analog weighing platforms and strain gauge load cells to a SysTec weighing terminal. It reads up to eight 350 Ω load cells and supplies them with 5 V excitation.
8Max. 350 Ω strain gauge load cells5VLoad cell excitation (±5 %)524,000dInternal resolution4 / 6wireLoad cell connectionConfirm your terminal model: The published ADM specifications differ between the WT1/WT3 and the WT4000/WT6000/WT8000 families (see below), and each family has its own calibration manual. Send us your terminal model and serial number so we can confirm the right module before you order.02 / 05Specifications by terminal familyThe ADM is specified separately for each family.
SysTec calls the terminals IT1, IT3, IT4000E, IT6000E and IT8000E; WeightechUSA sells them as WT1, WT3, WT4000, WT6000 and WT8000. These are the figures each family's manual gives for its ADM.
WT1 / WT3IT1 and IT3 manuals
W&M approved resolution 10,000 e. Update rate 50–800 updates per second (selectable in Service Mode). Calibration manual: IT1/IT3 ADM/DADM.
WT4000IT4000E manual
W&M approved resolution 6,000 e at a max. preload of 80 %. Update rate 50–400 updates per second. Calibration manual: ADM / DUAL-ADM / ADM8000-Exi.
WT6000IT6000E manual
W&M approved resolution 6,000 e at a max. preload of 80 %. Update rate 50–400 updates per second. Calibration manual: ADM / DUAL-ADM / ADM8000-Exi.
WT8000IT8000E manual
Same figures as the IT4000E and IT6000E; one or two scale interface modules can be installed. Calibration manual: ADM / DUAL-ADM / ADM8000-Exi.
03 / 05Load cell connectionWhat the ADM can read.
01Up to eight load cells
Max. 8 strain gauge load cells of 350 Ω each, overall impedance 43 to 3,300 Ω.
024-wire or 6-wire
Connects in 4-wire or 6-wire mode; the IT1 manual shows both wiring diagrams.
03Sealable calibration
Scale parameters are stored in EEPROM and secured with jumper W1, which can be sealed for approved (W&M) applications.
04 / 05Technical dataADM specifications.
Brand / model SysTec ADM (Analog Digital Module) SKU 20495 Function Scale interface for analog weighing platforms and strain gauge load cells Load cells Max. 8 strain gauge load cells, 350 ohm each Overall impedance 43 to 3,300 ohm Load cell excitation 5 V +/-5 % (gated supply) Connection 4-wire or 6-wire Internal resolution 524,000 d Min. input signal, approved use 0.33 uV / e Approved resolution, WT1 / WT3 10,000 e (update rate 50-800 per second) Approved resolution, WT4000 / WT6000 / WT8000 6,000 e at max. 80 % preload (update rate 50-400 per second) 05 / 05Part number and FAQOrder code.
Product Order code SysTec ADM analog digital module (A/D converter) 20495 Does one ADM fit every WT terminal?
We do not claim that. The manuals publish different ADM specifications for the WT1/WT3 and the WT4000/WT6000/WT8000 families, and list variants such as DADM and DUAL-ADM for two scales. Tell us your terminal model and we will confirm the module.
How many load cells can it read?
Up to eight 350 Ω strain gauge load cells (overall impedance 43 to 3,300 Ω).
Is it for analog or digital scales?
Analog. It connects analog weighing platforms and strain gauge load cells. Digital scale connections have their own calibration manuals.
Talk to a WeightechUSA specialist.
Questions about fit, calibration, certification, or volume pricing? Our team works on quotes the same day.
Email SalesCall +1 (954) 666-0877 × 305 Internal Battery -- Short name ADM Analog Digital Module